PMID: 8953435Nov 1, 1996Paper

A microstructural flow analysis within a bileaflet mechanical heart valve hinge

The Journal of Heart Valve Disease
J M GrossA P Yoganathan

Abstract

During recent clinical trials, the Medtronic Parallel bileaflet heart valve was found to have an unacceptable thrombosis complication rate. As patient- and material-related factors proved negative causes for this outcome, it was hypothesized that the flow fields within the valve's hinge pocket contributed to the thrombus formation. A microstructural flow analysis within the hinge pocket is presented which uses the techniques of flow visualization, computational fluid dynamics (CFD), and laser Doppler velocimetry (LDV). The application of these techniques towards solving this problem has become possible through (i) the ability to manufacture dimensionally correct 1-X transparent heart valve housings, (ii) advances in CFD technology, and (iii) advances in LDV measurement techniques. This analysis showed that a vortex was present at the hinge pocket's inflow channel during forward flow and degenerated to a disturbed three-dimensional structure during reverse flow with zones of turbulent shear stress large enough to cause blood cell damage. In addition, multiple zones of flow stagnation and disturbed flow existed along the leaflet's pivot throughout the entire cardiac cycle. It was felt that these complex fluid structures created c...Continue Reading