PMID: 20649004Jul 24, 2010Paper

A wireless power transmission system for capsule endoscope

Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi
Wenhui XinWenxing Wang

Abstract

In order to deliver power to the capsule endoscope, whose position and orientation are always changing when traveling along the alimentary tract, a wireless power transmission system based on electromagnetic coupling was proposed. The system is composed of Helmholtz transmitting coil and three-dimensional receiving coil. Helmholtz coil outside the body generates a uniform magnetic field covering the whole alimentary tract; three-dimensional coil inside retrieves stable power regardless of its position and orientation. The transmitter and receiver were designed and implemented, and the experiments validated the feasibility of the system. The results show that at least 320 mW of usable power can be transmitted to capsule endoscope when its position and orientation are changing at random and the transmitting power is 25W.

Related Concepts

Related Feeds

Cajal Bodies & Gems

Cajal bodies or coiled bodies are dense foci of coilin protein. Gemini of Cajal bodies, or gems, are microscopically similar to Cajal bodies. It is believed that Cajal bodies play important roles in RNA processing while gems assist the Cajal bodies. Find the latest research on Cajal bodies and gems here.