PMID: 29307159Jan 9, 2018Paper

Calculation of Mitral Valve Area in Mitral Stenosis: Comparison of Continuity Equation and Pressure Half Time With Two-Dimensional Planimetry in Patients With and Without Associated Aortic or Mitral Regurgitation or Atrial Fibrillation

Acta Medica Iranica
Roya SattarzadehBabak Geraiely

Abstract

Accurate measurement of Mitral Valve Area (MVA) is essential to determining the Mitral Stenosis (MS) severity and to achieving the best management strategies for this disease. The goal of the present study is to compare mitral valve area (MVA) measurement by Continuity Equation (CE) and Pressure Half-Time (PHT) methods with that of 2D-Planimetry (PL) in patients with moderate to severe mitral stenosis (MS). This comparison also was performed in subgroups of patients with significant Aortic Insufficiency (AI), Mitral Regurgitation (MR) and Atrial Fibrillation (AF). We studied 70 patients with moderate to severe MS who were referred to echocardiography clinic. MVA was determined by PL, CE and PHT methods. The agreement and correlations between MVA's obtained from various methods were determined by kappa index, Bland-Altman analysis, and linear regression analysis. The mean values for MVA calculated by CE was 0.81 cm (±0.27) and showed good correlation with those calculated by PL (0.95 cm, ±0.26 ) in whole population (r=0.771, P<0.001) and MR subgroup (r=0.763, P<0.001) and normal sinus rhythm and normal valve subgroups (r=0.858, P<0.001 and r=0.867, P<0.001, respectively). But CE methods didn't show any correlation in AF and AI s...Continue Reading

Related Concepts

Related Feeds

Atrial Filbrillation

Atrial fibrillation refers to the abnormal heart rhythm characterized by rapid and irregular beating of the atria. Here is the latest research.

Atrial Fibrillation

Atrial fibrillation is a common arrhythmia that is associated with substantial morbidity and mortality, particularly due to stroke and thromboembolism. Here is the latest research.

Arrhythmia

Arrhythmias are abnormalities in heart rhythms, which can be either too fast or too slow. They can result from abnormalities of the initiation of an impulse or impulse conduction or a combination of both. Here is the latest research on arrhythmias.