Cooling detection thresholds in the assessment of diabetic sensory polyneuropathy: comparison of CASE IV and Medoc instruments
Abstract
Cooling detection threshold testing may be an important quantitative method for assessing polyneuropathy, in that it has traditionally been viewed as a measure of small-fiber involvement. The present study sought to determine the agreement between two common testing devices and to determine whether these are concordant in their association with predictor variables for diabetic sensory polyneuropathy. A total of 83 patients with diabetes (10 patients with type 1 diabetes and 73 patients with type 2 diabetes) and a wide spectrum of diabetic sensory polyneuropathy severity underwent concurrent cooling detection threshold testing using the Medoc and CASE IV instruments. Common predictor variables for diabetic sensory polyneuropathy were measured on the same day. Measurements of cooling detection thresholds by both instruments were highly correlated (Spearman's correlation coefficient 0.81, P < 0.001) and demonstrated a high degree of agreement by the method of Bland and Altman (95% distribution critical values for the difference in cooling detection thresholds, +7.5 and -5.6 degrees C). Cooling detection thresholds by both instruments were strongly correlated with clinical indicators of large-fiber neuropathy but not with the sympt...Continue Reading