Design and In Vivo Verification of a CMOS Bone-Guided Cochlear Implant Microsystem

IEEE Transactions on Bio-medical Engineering
Xin-Hong QianChung-Yu Wu

Abstract

To develop and verify a CMOS bone-guided cochlear implant (BGCI) microsystem with electrodes placed on the bone surface of the cochlea and the outside of round window for treating high-frequency hearing loss. The BGCI microsystem consists of an external unit and an implanted unit. The external system-on-chip is designed to process acoustic signals through an acquisition circuit and an acoustic DSP processor to generate stimulation patterns and commands that are transmitted to the implanted unit through a 13.56 MHz wireless power and bidirectional data telemetry. In the wireless power telemetry, a voltage doubler/tripler (2X/3X) active rectifier is used to enhance the power conversion efficiency and generate 2 and 3 V output voltages. In the wireless data telemetry, phase-locked loop based binary phase-shift keying and load-shift keying modulators/demodulators are adopted for the downlink and uplink data through high-Q coils, respectively. The implanted chip with four-channel high-voltage-tolerant stimulator generates biphasic stimulation currents up to 800 μA. Electrical tests on the fabricated BGCI microsystem have been performed to verify the chip functions. The in vivo animal tests in guinea pigs have shown the evoked third ...Continue Reading