PMID: 9170130May 1, 1997Paper

Dual sensor pacemakers in children: what is the choice of sensor blending?

Pacing and Clinical Electrophysiology : PACE
Alpay CelikerSencan Ozme

Abstract

Dual sensor pacemakers were developed to obtain more appropriate responses to activity. We evaluated ten children with dual sensor pacemakers in different sensor blending circumstances using exercise testing to assess which ratio was optimal. Ten patients with several bradydysrhythmias (ages 6-16 years; mean 10.1 years) were included in the study. Eight patients had VVIR pacemakers (Vitatron Topaz), models and two patients had VDD pacemakers implanted via the transvenous route. All patients were in a paced rhythm (98.5% pacing). Accurate T wave sensing ranged from 81%-100%; mean 92%, median 95%. Voluntary exercise testing with a CAEP protocol was performed using a treadmill with the pacemaker in VVIR mode. Medium activity threshold with three sensor blending ratios (QT = ACT, QT > ACT, and QT < ACT) were done in all patients. The mean duration of exercise was not statistically different among the three sensor blending ratios. After 90 seconds of exercise, the mean pacing rate had increased by 12%, 3%, and 5%, respectively, in the three groups. At maximal exercise, the increases were 45%, 42%, and 54%. Mean HRs during exercise in each of the three ratios were not significantly different, although we found a statistically signifi...Continue Reading

References

Nov 1, 1992·Pacing and Clinical Electrophysiology : PACE·F M van KriekenM Sigmund
Nov 1, 1992·Pacing and Clinical Electrophysiology : PACE·F ProvenierL Jordaens
Dec 1, 1990·Pacing and Clinical Electrophysiology : PACE·M A LandmanN M van Hemel
Aug 1, 1988·Pacing and Clinical Electrophysiology : PACE·E AltH Blömer
Nov 1, 1988·Pacing and Clinical Electrophysiology : PACE·M ZegelmanJ Kreuzer
Jan 1, 1985·Pacing and Clinical Electrophysiology : PACE·D P HumenG J Klein
Apr 1, 1989·The American Journal of Cardiology·C P LauA J Camm
Jul 1, 1993·Pacing and Clinical Electrophysiology : PACE·M E CabreraJ Liebman
Jul 1, 1993·Pacing and Clinical Electrophysiology : PACE·R CowellA J Camm
Sep 1, 1993·Pacing and Clinical Electrophysiology : PACE·B S Goldman