Fate of the retained reservoir after replacement of 3-piece penile prosthesis
Abstract
We evaluated patients who underwent surgery for implant malfunction to determine whether retaining the reservoir was associated with an increased risk of erosion or infection. We also reviewed the literature to study possible risk factors for reservoir erosion. We reviewed the records of 85 patients who underwent 98 procedures for replacement of a malfunctioning 3-piece penile implant. Of these patients 13 underwent 2 replacement procedures. At the time of primary surgery the reservoir was placed in either the retropubic space or extraperitoneally in the lower lateral abdomen if the patient had undergone prior pelvic surgery. During surgery for replacement of the malfunctioning implant the cylinders and pump were removed, and the reservoir of the original 3-piece device was retained. During followup patients were assessed for implant function and the development of any complication such as infection and/or reservoir erosion into the bladder or bowel. Infection developed in 1 patient and implant malfunction occurred in 13 but no patient had erosion of the retained reservoir. All 85 patients had a functioning implant at a mean followup of 50 months (range 12 to 148). Review of the literature suggests that prior pelvic surgery and...Continue Reading
References
Citations
Intravesical penile implant reservoir: case report, literature review, and strategies for prevention
Related Concepts
Related Feeds
Bladder Carcinoma In Situ