PMID: 9181997Dec 1, 1996Paper

First experience in treatment of terminal cardiac insufficiency using multisite stimulation

Bulletin de l'Académie nationale de médecine
S CazeauJ Mugica

Abstract

We hypothesized that the presence of an abnormal ventricular mechanical activation sequence and/or a delayed left ventricular (LV) contraction may have adverse hemodynamic effects in congestive heart failure (CHF) and could be improved by synchronous RV-LV pacing in a multisite (MS) configuration. 8 NYHA IV CHF patients were included with a LV delay due to 1/ preexistent pacemaker in 4 pts (2 VVI and 2 DDD); 2/ left bundle branch block in 2 pts; 3/ intraventricular conduction delays in 2 pts. An acute hemodynamic evaluation was performed. Hemodynamics were optimized in standard RV pacing by modifying RV lead position from apex to outflow tract (RVOT) in VVI for AF patients and in VDD for sinus rhythm patients at different AV delays. RV pacing did not change hemodynamics whatever the lead position. BV pacing improved CI by 25% (p < 0.006), V wave by 26% (p < 0.004) and PCWP by 17% (p < 0.01). Chronic implantation was performed in 7pts. LV lead was implanted via the coronary sinus in 2 cases and epicardial via a thoracoscopic approach in the remaining ones. 1 pt died during LV lead implantation. Hemodynamics were tested at 2 months followup (FU). Switching BV pacing off was associated with immediate deterioration. At 6 +/- 6 mont...Continue Reading

Related Concepts

Related Feeds

Cardiac Conduction System

The cardiac conduction system is a specialized tract of myocardial cells responsible for maintaining normal cardiac rhythm. Discover the latest research on the cardiac conduction system here.