In-hospital experience with multiprogrammable implantable antitachycardia/antifibrillation devices

European Heart Journal
B WaldeckerH Tillmanns

Abstract

Multiprogrammable, automatic internal defibrillators with (n = 45) and without (n = 15) antitachycardia pacing features were implanted in 60 consecutive patients with refractory, malignant ventricular tachycardia (VT) (n = 42) or fibrillation (VF) (n = 18). Left ventricular (LV) ejection fraction was reduced to 39% +/- 12% as a result of structural heart disease in 56 patients. The complexity of the systems caused no additional risks to the surgical procedure or postoperative management. VT/VF detection parameters were individually adjusted to the arrhythmia type (detection cycle length 323 +/- 40 ms in patients with VF vs 405 +/- 40 ms for VT patients, P < 0.05) and incidence (longer detection periods if frequent nonsustained VT was also present). Shock energy was reduced in patients with VT as compared to VF (11J vs 24J, P < 0.05). Antitachycardia pacing was activated in 19/28 (68%) patients with well tolerated VT. Signal, telemetry, as detected by the device, combined with programmability allowed the device to be checked for correct decisions (these were inappropriate in four patients in three of whom corrections were non-invasive) prior to discharge. In conclusion, in the automatic tachyarrhythmia control devices we studied...Continue Reading

Citations

Mar 1, 1994·Pacing and Clinical Electrophysiology : PACE·H J TrappeB Kielblock
Apr 20, 2001·International Endodontic Journal·C A Murray, W P Saunders

❮ Previous
Next ❯

Related Concepts

Related Feeds

Atrial Fibrillation

Atrial fibrillation is a common arrhythmia that is associated with substantial morbidity and mortality, particularly due to stroke and thromboembolism. Here is the latest research.

Cardiology Journals

Discover the latest cardiology research in this collection of the top cardiology journals.

Arrhythmia

Arrhythmias are abnormalities in heart rhythms, which can be either too fast or too slow. They can result from abnormalities of the initiation of an impulse or impulse conduction or a combination of both. Here is the latest research on arrhythmias.