MO-D-BRB-08: BEST IN PHYSICS (THERAPY) - A Real Time Dose Monitoring and Dose Reconstruction Tool for Patient Specific VMAT QA and Delivery

Medical Physics
Kai YangNeelam Tyagi

Abstract

To develop a real time dose monitoring and dose reconstruction tool to identify and quantify sources of errors during patient specific VMAT delivery and QAMethods: The VMAT delivery monitor tool called Linac Data Monitor (LDM) has been developed that connects to the linac in clinical mode and displays, records and compares real-time machine parameters to the planned parameters. A new quantity called integral error keeps a running total of leaf overshoot and undershoots errors in each leaf pair multiplied by leaf width and the amount of time during which error exists in MU delivery. Another tool reconstructs pinnacle format delivered plan based on the saved machine logfile and recalculates actual delivered dose in patient anatomy. Delivery characteristics of various standard and hypofractionation VMAT plans delivered on Elekta Axesse and Synergy linacs were quantified. The MLC and gantry errors for all the treatment sites were 0.00±0.59mm and 0.05±0.31°, indicating a good MLC gain calibration. Standard fractionation plans had a larger gantry error than hypofractionation plans due to frequent dose rate changes. On average the MLC errors were negligible but larger errors of 4-6 mm and 2.5° were seen when dose rate varied frequentl...Continue Reading