MO-D-BRB-09: Treatment Delivery QA for Online Adaptive Radiation Therapy Based on Dynamic Machine Information (DMI): A Feasibility Study

Medical Physics
Taoran LiQ J Wu

Abstract

To implement a quality assurance (QA) system for the treatment delivery of online adaptive radiation therapy utilizing Dynamic Machine Information (DMI). DMI provides the expected/actual MLC leaf-positions, delivered MU, and beam-on status every 50ms during delivery. In this study a stream of DMI inputs is simulated by playing back Dynalog information recorded while delivering a test fluence map (FM). Based on these DMI inputs, the QA system performs three levels of monitoring/verification on the plan delivery process: (1) Following each input, actual and expected FMs delivered up to the current MLC position is dynamically updated using corresponding MLC positions in the DMI. The magnitude and frequency of pixel-by-pixel fluence differences between these two FMs are calculated and visualized in histograms.(2) At each control point, actual MLC positions are verified against the treatment plan for potential errors in data transfer between the treatment planning system (TPS) and the MLC controller.(3) Both (1) and (2) can signal beam-hold with a user-specified error tolerance.(4) After treatment, delivered dose is reconstructed in TPS based on DMI data during delivery, and compared to planned dose. (1) Efficiency: Average latency ...Continue Reading