Sensitivity and specificity of cerebrospinal fluid glucose measurement by an amperometric glucometer.

Saudi Medical Journal
Ahmed S AlkhalifahZainab S Alobaid

Abstract

To provide more data regarding the role of an amperometric glucometer in diagnosing meningitis. Methods: This is a prospective study conducted at the Pediatric and Neonatology Department, Qatif Central Hospital, Qatif, Saudi Arabia between March 2017 and September 2018. We measured glucose concentrations in cerebrospinal fluid (CSF) and blood using a central laboratory and amperometric glucometer (AG). We compared CSF/blood glucose ratios obtained in a central laboratory from clinical bedside examination with a glucometer, and calculated the sensitivity and specificity for detecting cases of meningitis. Results: A total of 101 patients with clinical suspicion of meningitis were recruited for CSF sampling. Of 101 CSF samples, 61 (60%) were suggestive of meningitis. Of 101 samples, 47 had hypoglycorrhachia identified by a standard laboratory, and 17% of them were also detected by AG. The correlation between CSF/blood glucose by AG and laboratory ratios was substantial (r=0.894, p  less than 0.01, 95% CI: 0.805-0.983). The AG sensitivity was 100% and specificity was 55% in pediatric cases, while in neonates the sensitivity was 86% and the specificity was 26%. Conclusion: Amperometric glucometers  can be used to detect hypoglycorrh...Continue Reading

Software Mentioned

Q Plot
SPSS
iSTAT CHEM8

Related Concepts

Related Feeds

CSF & Lymphatic System

This feed focuses on Cerebral Spinal Fluid (CSF) and the lymphatic system. Discover the latest papers using imaging techniques to track CSF outflow into the lymphatic system in animal models.

© 2022 Meta ULC. All rights reserved