SU-D-213AB-05: Commissioning a CT Compatible LDR T&O Applicator Using Analytical Calculation with ID and 3D Dosimetry

Medical Physics
J AdamsonO Craciunescu

Abstract

To determine the characteristics of a new commercially available CT-compatible LDR Tandem and Ovoid (T&O) applicator using 3D dosimetry. We characterized source attenuation through the asymmetric gold shielding in the buckets by measuring dose with diode and 3D dosimetry and compared to an analytical line integral calculation. For 3D dosimetry, a cylindrical PRESAGE dosimeter (9.5cm diameter, 9.2cm height) with a central 6mm channel bored for source placement was scanned with the Duke Large field of view Optical CT-Scanner (DLOS) before and after delivering a nominal 7.7Gy at a distance of 1 cm using a Cs-137 source loaded in the bucket. The optical CT scan time lasted approximately 15 minutes during which 720 projections were acquired at 0.5° increments, anda 3D dose distribution was reconstructed with a 0.5mm(3) isotropic voxel size. The 3D dose distribution was applied to a CT-based T&O implant to determine effect of ovoid shielding on the dose delivered to ICRU 38 Point A as well as D2cc of the bladder, rectum, bowel, and sigmoid. Dose transmission through the gold shielding at a radial distance of 1-3cm from midplane of the source was 86.6%, 86.1, and 87.0% for analytical calculation, diode, and 3D dosimetry, respectively....Continue Reading