SU-D-217A-06: Impact of Anterior-Posterior (AP) and Posterior-Anterior (PA) Scout Scans on the CT Radiation Dose in the Whole Body PET/CT Scan

Medical Physics
Dershan Luo, Tinsu Pan

Abstract

CT can contribute over 50% of radiation dose in the whole body (WB) PET/CT scan. Tube current modulation (TCM) is a standard technique for reducing CT radiation dose to the patient by changing the tube current with the patient size, and is controlled by a very low-dose scoutscan, which assumes the patient is positioned at the center of the CT gantry opening. However, most patients are not positioned at the center due to practicality or to avoid claustrophobic or to reduce time of radiation exposure from the patient to the technologist. We study the impact of the AP and PA scout scans to the patient radiation exposure from CT. Ina retrospective study of 200 patients, each received two WB PET/CT scans: one with AP, and the other one with PA. The helical CT with TCM and PET acquisitions were identical in both scans. Separation of the two scans was about 10 months in average. The scans were performed on four GE PET/CT scanners: three 16- and one 64-slice with the same TCM settings. The 200patients were selected for the same scan coverage and similar body weight (difference = 3 kg). The tube current in each slice and average exposure tothe patient were recorded and compared. The AP scout caused lower radiation dose on 94% of the pat...Continue Reading

Related Concepts

Related Feeds

Aphasia

Aphasia affects the ability to process language, including formulation and comprehension of language and speech, as well as the ability to read or write. Here is the latest research on aphasia.