SU-D-BRA-03: Simultaneous MV-KV Imaging for Intra-Fractional Motion Management during Volume Modulated Arc Therapy (VMAT) Delivery on the Varian TrueBeam

Medical Physics
Margie A HuntGig Mageras

Abstract

To evaluate a MV-kV intra-fractional imaging technique for use during volume modulated arc therapy (VMAT) with the Varian TrueBeam. MV-kV image pairs were acquired intra-fractionally during VMAT delivery. kV images (11 fps) were acquired throughout delivery using a standard pre-programmed imaging template. MV images (9.5 fps) were acquired simultaneously by deploying the EPID and passively collecting the resulting images using Varian proprietary software, iTools Capture. Localization accuracy was evaluated by imaging a Rando phantom implanted with 3 fiducials while moving the couch according to XML- programmed trajectories simulating typical prostate and respiratory motion. VMAT delivery was done using a single 360 degree arc in TrueBeam Developer mode. The effect on accuracy of total MU and gantry speed was studied. To improve image quality, MV frame averaging was performed and the MV and kV images were then registered to their corresponding DRRs using in-house registration software. From these 2D registrations, the 3D position at each MV-kV acquisition point was determined. Between 130 and 390 MV-kV pairs were acquired for each delivery. The mean difference between planned couch and measured fiducial 3D positions with prostat...Continue Reading