SU-E-T-164: Clinical Implementation of ASi EPID Panels for QA of IMRT/VMAT Plans

Medical Physics
K HosierS Rosenthal

Abstract

To investigate various issues for clinical implementation of aSi EPID panels for IMRT/VMAT QA. Six linacs are used in our clinic for EPID-based plan QA; two Varian Truebeams, two Varian 2100 series, two Elekta Infiniti series. Multiple corrections must be accounted for in the calibration of each panel for dosimetric use. Varian aSi panels are calibrated with standard dark field, flood field, and 40×40 diagonal profile for beam profile correction. Additional corrections to account for off-axis and support arm backscatter are needed for larger field sizes. Since Elekta iViewGT system does not export gantry angle with images, a third-party inclinometer must be physically mounted to back of linac gantry and synchronized with data acquisition via iViewGT PC clock. A T/2 offset correctly correlates image and gantry angle for arc plans due to iView image time stamp at the end of data acquisition for each image. For both Varian and Elekta panels, a 5 MU 10×10 calibration field is used to account for the nonlinear MU to dose response at higher energies. Acquired EPID images are deconvolved via a high pass filter in Fourier space and resultant fluence maps are used to reconstruct a 3D dose 'delivered' to patient using DosimetryCheck. Res...Continue Reading

Related Concepts

Related Feeds

Buruli Ulcer

Buruli ulcer is a progressive disease of subcutaneous tissues caused by mycobacterium ulcerans. Here is the latest research.

Related Papers

Technology in Cancer Research & Treatment
Luis A Vazquez QuinoNiko Papanikolaou
Journal of Applied Clinical Medical Physics
Dewayne L DefoorSotirios Stathakis
© 2022 Meta ULC. All rights reserved