SU-E-T-387: Validation of a New System for Patient Specific IMRT QA and Comparison with Other Commerical Systems

Medical Physics
R McKinseyNikos Papanikolaou

Abstract

The focus of this project is to compare the Octavius 4D with current commercial available dose validation systems: MatriXX MultiCube and Delta4. Many challenges are faced with properly measuring Intensity Modulated Radiotherapy (IMRT). It has become common practice for clinics to use film, arrays, or multiple detectors to validate dose measurements pretreatment for static and dynamic treatments. IMRT QAs for various treatment sites were measured for patients using three different dose validation systems. All measurements were taken on a Varian CLinac 2100 C/D, SN-757, 80 MLC with 6MV. The treatment plans evaluated were Step-N-Shoot. Data analysis was performed using the software provided with each dose validation system. Detailed information was gathered from each system with their perspective advantages. The latest system, Octavius 4D, allows one to calculate the Gamma Index for Coronal, Sagittal, and Transversal views for every slice included in the measurement along with the traditional data analysis provided; histograms, horizontal and vertical profiles, DTA. The Gamma Index values were observed using the MatriXX Multicube, Delta4, and Octavius 4D. The treatment plan included five fields at various gantry angles. Also the g...Continue Reading