SU-E-T-531: Verification of Acuros Dose Calculation Accuracy in Lung SBRT

Medical Physics
L WangEd Mok

Abstract

The AcurosXB Advanced Dose algorithm (Varian Medical Systems) was reported to give more accurate dose calculation in heterogeneous medium compared with the AAA algorithm. In this work, a validation of AcurosXB dose calculation accuracy is performed for use in lung SBRT. A CIRS (Computerized Imaging Reference Systems, Inc.) lung phantom was used with a tissue equivalent 2cm circular target inserted into the lung. The phantom was customized to hold a Gafchromic EBT2 film through the center of the target. The phantom was CT scanned, and single field (3cm by 3cm anterior/lateral) plans, as well as VMAT plans were created for 6 and 10 MV energies with different delivery modalities (with/without a flattening filter). All plans delivered 500 cGy to the PTV which was the target plus 0.5 cm uniform margin. Plans were delivered on a Varian TrueBeamTM STx using image guidance to locate the isocenter. The film dose was compared with Eclipse v10.0 dose calculated with AAA and Acuros algorithms using FilmQA software (3cognition). Monte Carlo simulations (BEAMnrc/DOSXYZnrc) were performed to further validate AcurosXB calculations. Within the GTV, AAA and Acuros algorithms were in good agreement with the film measurements to within 2% mean dos...Continue Reading

Related Concepts

Related Feeds

Aortic Aneurysm

An aortic aneurysm is the weakening and bulging of the blood vessel wall in the aorta. This causes dilatation of the aorta, which is usually asymptomatic but carries the risk of rupture and hemorrhage. Find the latest research on aortic aneurysms here.

Aneurysm

Aneurysms are outward distensions or bulges that occurs in a weakened wall of blood vessels. Discover the latest research on aneurysms here.