T wave oversensing in subcutaneous implantable cardioverter defibrillator secondary to hematoma formation: A potential cause of early postimplantation inappropriate shocks

Journal of Arrhythmia
Ahmad Halawa, Sandeep Gautam

Abstract

T wave oversensing (TWOS) is the most common cause of inappropriate shocks in subcutaneous cardioverter-defibrillators (S-ICD). We are presenting a patient with severe ischemic cardiomyopathy who received a S-ICD while on antiplatelets therapy. Pressure dressing was applied due to significant bleeding. On the first postoperative day, the device delivered 26 inappropriate shocks after removal of the pressure dressing. Interrogation revealed new TWOS, likely related to changes in the sensing vectors after hematoma formation.

Methods Mentioned

BETA
amputation

Software Mentioned

SMART

Related Concepts

Related Feeds

Cardiomyopathy

Cardiomyopathy is a disease of the heart muscle, that can lead to muscular or electrical dysfunction of the heart. It is often an irreversible disease that is associated with a poor prognosis. There are different causes and classifications of cardiomyopathies. Here are the latest discoveries pertaining to this disease.