The Systematic Bias of Ingestible Core Temperature Sensors Requires a Correction by Linear Regression
Abstract
An accurate measure of core body temperature is critical for monitoring individuals, groups and teams undertaking physical activity in situations of high heat stress or prolonged cold exposure. This study examined the range in systematic bias of ingestible temperature sensors compared to a certified and traceable reference thermometer. A total of 119 ingestible temperature sensors were immersed in a circulated water bath at five water temperatures (TEMP A: 35.12 ± 0.60°C, TEMP B: 37.33 ± 0.56°C, TEMP C: 39.48 ± 0.73°C, TEMP D: 41.58 ± 0.97°C, and TEMP E: 43.47 ± 1.07°C) along with a certified traceable reference thermometer. Thirteen sensors (10.9%) demonstrated a systematic bias > ±0.1°C, of which 4 (3.3%) were > ± 0.5°C. Limits of agreement (95%) indicated that systematic bias would likely fall in the range of -0.14 to 0.26°C, highlighting that it is possible for temperatures measured between sensors to differ by more than 0.4°C. The proportion of sensors with systematic bias > ±0.1°C (10.9%) confirms that ingestible temperature sensors require correction to ensure their accuracy. An individualized linear correction achieved a mean systematic bias of 0.00°C, and limits of agreement (95%) to 0.00-0.00°C, with 100% of sensors a...Continue Reading
References
Validity and reliability of devices that assess body temperature during indoor exercise in the heat.
Citations
Software Mentioned
Related Concepts
Related Feeds
Acute viral rhinopharyngitis
Acute viral rhinopharyngitis, also known as "common cold", is an acute, self-limiting viral infection of the upper respiratory tract involving the nose, sinuses, pharynx and larynx. Discover the latest research on acute viral rhinopharyngitis here.